EN – Operating Manual
CALIBRATION
To maintain the accuracy of the breathalyzer, ACE
Instruments recommends a calibration after 6 up to
maximum 12 months or after 1,000 tests (depending which
incident happens first). For a calibration service you can
send the device to ACE Instruments (see www.alkomat.net).
The calibration is not mandatory, but the accuracy of the
breathalyzer will decrease steadily without it and additionally
the warranty for the sensor expires.
We recommend sending it as a parcel, so that in case of
loss an inquiry or proof of delivery is possible.
BATTERY NOTICES
The device is run with three (3) AAA Alkaline batteries (1.5
V). These batteries are not made for recharging. The use of
a rechargeable batteries is not possible with this device and
can damage it. If the batteries are depleted, they must not
be thrown into the household waste. Used batteries may
contain harmful substances that can cause environmental
or health damage. Please return all depleted batteries to
electrical appliance stores, who collect them for free.

BATTERY EXCHANGE

To replace the batteries, push the battery cover on the back
of the unit downwards and replace the discharged batteries
with new ones. Be aware of the polarity when inserting the
batteries (+/- pole). Close the battery compartment cover
again. In the EU the return of used batteries is mandatory by
law. All depleted batteries and also rechargeable batteries
will be recycled. This way valuable resources like iron, zinc
and nickel can be retrieved and are reusable then. Battery
recycling is the easiest lesson in environmental protection –
so please support it!
The shown garbage bins mean that batteries or recharge-
able batteries must not be thrown into the household waste.
The signs below the garbage bins are acronyms for the
chemical substances inside the battery:
29
Hg
Cd
Hg: Battery contains Mercury
Cd: Battery contains Cadmium
Pb: Battery contains Lead

TECHNICAL DATA*

Sensor
Fuel Cell Sensor (Reference standard
complies with EN16280)
Accuracy
+/- 0.06‰ BAC (per mill) with
reference to 0.50‰ BAC at 25°C
(per mill = default setting)
OR +/- 0.006% BAC (Blood Alcohol
Concentration) with reference to
0.050% BAC at 25°C
OR +/- 0.03mg/L with reference to
0.250mg/L at 25°C
Detection
0.00 up to 4.00‰ BAC (per mill)
OR 0.000 up to 0.400% BAC (USA)
Range
OR 0.000 up to 2.000mg/L (UK)
Warm-up
Within 10 up to 15 seconds at 0.5‰ BAC,
Time
it may vary depending on the value of
per mill / BAC which was measured
Dimension
132.5 x 68.5 x 26mm (L x W x H)
87g (122g including 3 batteries)
Weight
Power
3 pieces of 1.5V AAA Alkaline battery
Supply
0°C up to 50°C
Working
Temperature
Pb
30