Summary of Settings and Options
The options for the various Omnipod DASH Insulin Management System
settings are:
Time format
Time zones
Date format
Screen time-out
PIN
Maximum Basal Rate
Basal rate
Basal Programmes
Basal rate segments
Temp basal
Temp basal (set to %)
Temp basal (set to U/hr)
Temp basal presets
BG Goal Range for blood
glucose history
BG reminder
Custom reminder
Bolus Calculator
Target BG value
Correct Above threshold
12-hour or 24-hour clock.
GMT-11:00 to GMT+13.00.
MM/DD/YY
DD/MM/YY
MM.DD.YY
DD.MM.YY
YY-MM-DD
30, 60, 120 seconds. Default is 30 seconds.
4 digits from 0 to 9.
0.05 to 30 U/hr. Default is 3.00 U/hr.
Units/hr. Range: 0 U/hr to Maximum Basal Rate in
0.05 U/hr increments.
Maximum of 12.
24 per Basal Programme.
%, units/hr or Off. Default is Off.
Duration: 30 min to 12 hrs in 30-min increments.
Range: 100% decrease (0 U/hr) to 95% increase from
current basal rate in 5% increments. Cannot exceed the
Maximum Basal Rate.
Range: 0 U/hr to Maximum Basal Rate in 0.05 U/hr
increments.
Maximum of 12.
Lower and upper limits: 3.9 to 11.1 mmol/L in
0.1 mmol/L increments.
On or Off. Default is Off.
Maximum of 4 active at one time.
Reminder can occur between 30 min to 4 hrs after bolus
is started. Set in 30-minute increments.
Maximum of 4. Set to Daily, Once only, Off.
On or Off. Default is On.
Maximum of 8 segments; 3.9 to 11.1 mmol/L in
0.1 mmol/L increments.
Maximum of 8 segments; Target BG to 3.9 to
11.1 mmol/L in 0.1 mmol/L increments.