Cable Junction Box MP 90/04
5.5

Electronic corner correction

5.5.1

General information

Minebea Intec
Minebea Intec load cells are produced according to high quality standards and have
precisely adjusted output values.
Nonetheless, mechanical imbalances can cause impermissible corner load errors to arise,
which will need to be offset by soldering in resistors.
The correct installation and accurate alignment of load cells are imperative for good
measurement results and significantly affect the behavior with corner loads. Therefore,
the installation and alignment of the load cell should always be checked first if a corner
error is identified.
If necessary, carry out mechanical height adjustment (see Installation manual of the load
cell).