Cable Junction Box MP 90/04

4 Installation and connection information

4.1

General information

4.2

Terminals and jumpers

Minebea Intec
-
Only use the cable glands provided by the manufacturer.
-
Install the junction box so that the cable glands are on the bottom.
-
Do not open the junction box when connected to the voltage supply.
Note:
At ambient temperatures >45°C, cables that are suitable for at least 85°C must be used.
Signal +
green
Signal -
gray
Supply -
blue
Sense -
black
Terminal contacts
+ measuring voltage (load cells)
1–4
Terminal contact
+ measuring voltage (connection cable)
IND
Terminal contacts
- measuring voltage (load cells)
1–4
Terminal contact
- measuring voltage (connection cable)
IND
Terminal contacts
- supply voltage (load cells)
1–4
Terminal contact
- supply voltage (connection cable)
IND
Terminal contacts
- sense (load cells)
1–4