Cable Gland - Minebea Intec Mp 90/04 Manuel D'installation

Masquer les pouces Voir aussi pour MP 90/04:
Table des Matières
Les langues disponibles

Les langues disponibles

4 Installation and connection information
4.3

Cable gland

EN-11
Sense +
white
Supply +
red
Screen
yellow
1
2
The cable glands each have one sealing ring.
The M20 cable gland can be fitted with a reduction sealing ring if cables ≤8 mm are used
(see Chapter 4.5).
The M25 cable gland must be fitted with a four-seal insert to be able to feed in up to four
load cell cables.
The cable wires are connected to the terminals inside the device.
NOTICE
Property damage is possible.
If a seal insert is not used, it must be sealed with a supplied locking pin.
NOTICE
Property damage is possible.
Regularly check the fitted cable gland for tightness and re-tighten it, if necessary.
Terminal contact
- sense (connection cable)
IND
Terminal contacts
+ sense (load cells)
1–4
Terminal contact
+ sense (connection cable)
IND
Terminal contacts
+ supply voltage (load cells)
1–4
Terminal contact
+ supply voltage (connection cable)
IND
Terminal contacts
Screen (load cells)
1–4
Terminal contact
Screen (connection cable)
IND
Jumper
Connects the load cell supply line to the
sense line. Both jumpers must be opened
Jumper
when 6-wire load cell cables are used.
Cable Junction Box MP 90/04
Minebea Intec