VISIOMED BewellConnect MyTens Pro BW-TS1 Mode D'emploi page 13

Masquer les pouces Voir aussi pour BewellConnect MyTens Pro BW-TS1:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

you should to perform this program at rest.
Summary Table of the TENS Programs
Treatment
Indication
site
Shoulder
pain and
Shoulder
frozen
shoulder
Pain in
Lumbar
the lower
areas
back
Pain in the
Elbow
elbow
Pain in the
Forearm
forearm
Pain in
the wrist
Wrist
and carpal
tunnel
syndrome
Pain in
Hip
the hip
Thigh
Sciatic pain
TENS
Treatment
Treatment
time
phases
Phase 1
Phase 2
30mn
Phase 3
Phase 1
Phase 2
Phase 3
30mn
Phase 4
Phase 5
Phase 6
Phase 1
30mn
Phase 2
Phase 1
Phase 2
30mn
Phase 3
30mn
Phase 1
Phase 1
Phase 2
Phase 3
30mn
Phase 4
Phase 5
Phase 6
30mn
Phase 1
BewellConnect - BW-TS1 - User Manual - 122016V1
Impulses
Frequency
Width
(time)
in µS
2Hz (10s)
250
4Hz (8s)
250
6Hz (6s)
250
2Hz (10s)
250
4Hz (8s)
250
6Hz (6s)
250
2Hz (10s)
250
4Hz (8s)
250
6Hz (6s)
250
2Hz (3s)
150
100Hz (3s)
200
2Hz (3s)
200
100Hz (3s)
150
3Hz
250
2Hz
250
6Hz (30s)
250
6Hz (30s)
250
8Hz (20s)
250
8Hz (20s)
250
10Hz (20s)
250
10Hz (20s)
250
100Hz
200
Wave form
Description
Continuous
CONT
output
Continuous
CONT
output
Alternating
HANS
output
HANS
Alternating
output
HANS
Up : 1.5s
Down : 3s
EMS
10s cycle
with 2/5
strength
Continuous
CONT
output
Continuous
CONT
output
Up : 0.5s
Cont : 7s
EMS
Down : 0.5s
Rest : 7s
13

Publicité

Chapitres

Table des Matières
loading

Table des Matières