BTL7-A/C/E5 _ _ -M _ _ _ _ -SF-F _ _
Magnetostrictive Linear Position Sensor – Rod Style
6
Calibration procedure (continued)
6.3.2

Adjust

The detailed procedure for adjusting is
described on page 16 ff.
The factory-set null point and end point are replaced by a
new start point and end point and the associated output
values can be adjusted. The start and end values can be
adjusted as you like up to the limits.
Steps:
► Move float to the new start position.
► Read new start point by activating the programming
inputs.
► Set the desired start value by activating the
programming inputs.
New start
value
New null point
Fig. 6-4:
Adjust a new start value
14
english
► Move float to the new end position.
► Read new end point by activating the programming
inputs.
► Set the desired end value by activating the
programming inputs.
New end value
Before
Fig. 6-5:
Adjust a new end value
After
6.3.3

Reset

Restoring the BTL to its factory settings.
The detailed procedure for the reset is
described on page 18.
Before
After
New end point