TECHNICAL DESCRIPTION

1. DESCRIPTION

The Optilux 501 is a visible curing unit intended for polymerization of light
cured materials by dental professionals.

2. ERGONOMIC FEATURES

Many ergonomic features are incorporated into the Optilux 501 as defined
below:
a) Carrying handle:
The recessed back at the top of the unit was
designed to be used as a carrying handle. The unit is well balanced and,
with the handpiece in place, can be easily carried by placing fingers in
the recess and gripping the top of the unit.
b) Ventilation openings: The openings are hidden on the top and bottom
of the control unit to help avoid liquids from entering. Do not cover
openings.
c) Handpiece hanger: The hanger is designed to secure the handpiece
and prevent it from becoming accidentally dislodged when the unit is
carried.
In addition, the hanger material is durable and resistant to
breakage.
d) Handpiece placement: The handpiece can be placed in the hanger,
pointing in either direction for left or right handed operators, without
compromising the secure fit.
e) Easy grip:
With the handpiece resting in its hanger, the handle is
positioned away from the front surface of the control unit so that a gloved
hand can slip around the handle and grip it comfortably. Remove the
handpiece by rotating the handle forward while lifting the handpiece out
of the hanger.
Keypad: The raised keys of the keypad provide a tactile feel when
f)
depressed, with audio and visual confirmation of the selection.
g) Smooth equipment surfaces: The Optilux is designed with smooth
surfaces for easy cleaning.
DISINFECTING/STERILIZING for details.
Refer to SECTION 11, CLEANING/

3. SPECIFICATIONS AND EQUIPMENT CLASS

a)
AC supply connection:
b)
Power input:
c)
Equipment class:
d)
Protection from
electric shock:
e)
Protection from
ingress of liquids:
f)
Environment:
g)
Transport and storage:
Ambient temperature
Relative humidity
Atmospheric pressure
h)
Operation:
i)
Fuses (2 per unit):
j)
Lamp:
k)
Output wavelength range:
l)
Output light Intensity:
• 10, 20, 30, 40, C
• R, B, BLEACH
m)
Curing radiometer:
Sensitivity
Intensity range
Accuracy
n)
Standard fiber optic light guide:
Accessory fiber optic light guide:
o)
Unit overall dimensions:
Height
Length
Width
p)
Unit weight:
Handpiece
Control Box
q)
Thermal safety:
r)
Additional symbols:
Attention, consult
accompanying
documents
Alternating
Current
2
100-120V~/2.0A, 50/60 Hz
220-240V~/1.0A, 50/60 Hz
160 VA, nominal
II
Type BF
None
Equipment not suitable for use in the presence
of a flammable anaesthetic mixture with air or
nitrous oxide
-40°F to 158°F (-40°C to +70°C)
10% to 100%, including condensation
.5 atm to 1.0 atm (500 hPa to 1060 hPa)
Continuous operation with intermittent loading,
duty cycle of 1 minute ON/4 minutes OFF
100-120V, T 1.6 A, part no. 30257
220-240V, T 0.8 A, part no. 30246
12 Volt, 80 Watt OptiBulb, part no. 21237
400-505 nm (nanometers)
850 mW/cm² minimum using 8mm curved turbo+ light guide
1000 mW/cm² minimum using 8mm curved turbo+ light guide
(mW/cm² = milliwatts per centimeter square)
400-510 nm
100-1999 mW/cm²
100 mW/cm²
8mm turbo+ curved light guide, part no. 952213
11mm curved light guide, part no. 20898
8.0" (20.5 cm)
7.5" (19.0 cm)
6.25" (16.0 cm)
10 oz. (300 g.)
6 lb. 6 oz. (2.9 kg.)
Handpiece thermostat protection
"OFF"
"ON"