4.
TROUBLESHOOTING
4.1
Connecting and Disconnecting the Printer
The printer is automatically detected and enabled when connected to the trophon EPR.
To enable or disable the printer manually, use the trophon EPR screen by selecting Menu
Enable/Disable.
Printer – Error Messages
4.2
Printer error messages are displayed on the screen of the trophon EPR. If an error message is
displayed, two options are available:
1. Retry - select via the trophon EPR soft keys
 After the error has been addressed, select Retry
 The trophon will attempt to re-print the last disinfection record. Check the printed record
before proceeding to remove the probe from the trophon EPR chamber.
 If the last disinfection record did not reprint, the two options Retry and Ignore will again
display.
2. Ignore – select via the trophon EPR soft keys
 If the printed record is NOT required and you do NOT want to Retry the print, select Ignore.
 After Ignore has been selected, remove the probe from the trophon EPR Chamber.
NOTE: If Ignore is selected, the disinfection label record will not be printed.