Measurement and instrumentation
UT51 multimeter
Ref :
291016
ENGLISH
All manuals and user guides at all-guides.com
Overload protection:
0.3 A / 250 V fuse
o
10 A / 250 V fuse
o
4.4 AC current
Range of values:
200 µA, 2 mA, 20 mA, 200 mA, 2 A, 10 A
o
Resolution:
0.1 µA, 1 µA, 10 µA, 100 µA, 1 mA, 10 mA
o
Accuracy:
± (1.8 % of the displayed value + 3 digits) for the 200 µA
o
range
± (1 % of the displayed value + 3 digits) for the 2 mA and 20
o
mA ranges
± (1.8% of the displayed value +3 digit) for the 200 mA and 2
o
A ranges
± (3 % of the displayed value +7 digits) for the 10 A range
o
Overload protection:
0.3 A / 250 V fuse
o
10 A / 250 V fuse
o
Display:
RMS value of the sine wave (average value)
o
4.5 Resistance
Ranges of values:
200 Ω, 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 20 MΩ
o
Resolution:
Ω, 1 Ω, 10 Ω, 100 Ω, 1 kΩ, 10 kΩ
o
Accuracy:
± (0.8 % of the displayed value + 5 digits) on the 200 range
o
± (0.8 % of the displayed value + 2 digits) for the 2 k to 2 M
o
range
± (1 % of the displayed value + 5 digits) for the 20 M range
o
Overload protection:
230 V (DC or AC)
o