Connecting Transmitter to Energized working system
The Transmitter, with the red test lead, can be directly connected
to the live wire of the working electrical equipment under load
(motor, electronics, etc). Tracing can be performed without needing
to turn off the equipment or switching power off.
Proper connection with separate ground
Alternately in some cases a separate ground connection could be
used instead of separate neutral. In this case please connect the
green wire to a separate ground wire / Protective Earth (PE) on a
separate circuit instead.*
*Note: Please note that if working with RCD protected circuits,
separate ground connection will trip the RCD. Refer to Special
Applications, section 4.1 "RCD-Protected Circuit Wire Tracing" for
alternative tracing methods in manual of AT-6010.
ATTENTION: Due to safety reasons in energized circuits the separate
ground connection must always be connected to a well connected
Protective Earth (PE) of an electrical installation.
Only for de-energized circuits the separate ground could be also
connected alternative to the metal building structure or metal
water pipe.
©2019 Beha-Amprobe
For detailed specifications and ordering go to beha-amprobe.com 6011881 A
®
Advanced Wire Tracer - Quick Guide