Data logging
CO
probe
2
Ref :
453 076
ENGLISH
6- Select Air by pressing the orange button.
7- Press the setup button (violet button) at least for 2 seconds. A triangle
blinks above the pictogram
8- When the arrows blink alternately
Probe in the air, set 0 by using the potentiometer
violet button
: a triangle blinks above the pictogram
9- Making of the calibration solution : pour 20 ml of base solution into a flask.
Add 3 ml of acid solution.
10- Put the probe into the calibration solution.
The displayed value remains the same.
Caution : the lifetime of the solution is 10 to 15 min. If exceeded, make
a new solution.
11- When the arrows blink alternately
Turn the potentiometer
setup button
.
12- If the setup is OK a triangle appears above the pictogram « OK ».
In case of probe failure, a triangle appears above the pictogram
The probe is ready to be used in Air or Water !
17
.
, the probe is ready to be setup.
, then briefly press the
, the probe is ready to be setup.
until the value 5,8 is displayez then press the
.
.