2. Brush off internal and external contaminants:
– Insert the cleaning brush into the proximal argon gas
connection as far as it will go and pull out again. Clean
the brush if necessary.
– Then repeat the brushing step at least twice, thereby re-
moving as much contamination as possible.
– Switch to the distal instrument connection.
– Insert the cleaning brush into the instrument connection
as far as it will go and pull out again. Clean the brush if
necessary.
– Then repeat the brushing step at least twice, thereby re-
moving as much contamination as possible.
– Lastly, brush off the outer surface of the product (Note:
Use a short handle brush).
3. Flush out and rinse the product with warm water to remove
any contaminants and cleaning agent residue:
– Screw the luer-lock adapter back onto the argon gas
connection of the handle.
– Use a syringe to flush out the internal lumen until the
water runs clear.
– Thoroughly rinse the outer surface of the product.
4. Flush out and rinse the product with sterile water. Proceed
as in the previous step.
5. Remove the syringe and luer-lock adapter.
6. Dry the inside and outside of the product:
– Use a syringe or compressed air (max. 29 psi) to blow
out the internal lumen until no more liquid is discharged.
– Completely dry the outer surfaces of the product.
7. If contaminants are still visible on the product or come out of
the product after cleaning/disinfection, repeat the entire
process.
40