If damaged, do not use this product!
Do not use in the presence of combustible or explosive materi-
als!
Never lay this product on the patient or in his/her direct vicinity.
Damp inside the connectors may cause flash-overs that could
damage the connector or its insulation. Ensure that the product
is dried sufficiently during processing and prevent fluids from
penetrating the connector during use. The older the material, the
greater the risk of unwanted damp in the connectors.
Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.
4
How to use
Clean, disinfect and sterilize the product before using it for the
first time and before every subsequent use.
WARNING! Do not kink the cable.
WARNING! Protect the product from all physical damage! Do not
throw! Do not apply force of any kind!
WARNING! Risk of injury! Do not activate the electrosurgical unit
while you are connecting the cable.
Connect the instrument to the cable. Then connect the connect-
ing cable to the unit's APC socket as follows:
• Insert the HF connector into the multifunctional socket. The
connector must click into place audibly.
• Screw the argon gas connector to the argon gas socket. Erbe
recommends using the membrane filter (art. no. 20132-059)
to protect both patient and unit.
21