OBJ_DOKU-189884-007.fm Seite 28 Freitag, 11. Juni 2021 10:59 10
• Machine cleaning and disinfection in a G 7836 CD disinfector
(Miele & Cie. KG, Gütersloh, Germany) with the cleaning
agent neodisher mediclean forte (Dr. Weigert GmbH & Co.
KG, Hamburg, Germany).
– neodisher mediclean forte: Dosage 0.5%, 5 minutes ex-
posure time at 55 °C
– Disinfection: 5 minutes at 90 °C.
• Sterilization in an autoclave Systec V-150 (Systec GmbH,
Linden/Germany).
– Steam sterilization with saturated steam, fractionated
vacuum procedure, 3 minutes, 132 °C, 3 bar.
Erbe recommends the processing procedures described in these
instructions for use. Equivalent different procedures are possible
if not explicitly excluded. It is incumbent on the user to ensure
the suitability of the actual procedures used by means of suit-
able measures (e.g. validation, routine monitoring, check of ma-
terial compatibility).
6
Disposal
Dispose of the product, packaging material and accessories (if
available) in accordance with the national guidelines and legis-
lation applicable in each case.
7
Symbols
Symbol
Explanation
Consult instruc-
tions for use
Catalogue number
Symbol
Explanation
Caution, consult
accompanying
documents
Batch code