Cleaning, Disinfection, Sterilization
5.1
Safety instructions for processing
Erbe Elektromedizin recommends machine cleaning/
disinfection.
May be cleaned in an ultrasonic bath.
On no account use sharp or abrasive objects for cleaning purpos-
es.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 29 psi (2 bar).
Maximum water pressure during rinsing: 29 psi (2 bar).
During machine cleaning and disinfection, do not exceed 203 °F
(95 °C).
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of Validated
Procedures".
During sterilization, do not exceed 280 °F (138 °C).
Do not sterilize in hot air oven.
5.2
Processing Limitation
Frequent processing can damage this product. Observe the safe-
ty instructions for checking the product. With visible damage or
functional impairment, the product may no longer be used.
5.3
Disassembly
1. Detach the instrument from the cable without damaging in-
sulation of the cable or instrument.