Télécharger Imprimer la page

R82 High-low Mode D'emploi page 34

Masquer les pouces Voir aussi pour High-low:

Publicité

Test modes during immunity tests:
Battery mode, standby armed/ready.
Charging mode, actuator not running.
EM immunity standards
Electrostatic discharge
EN
IEC61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields form RF wireless
communication equipment
IEC 61000-4-3
Electrical fast transients / bursts
IEC 61000-4-4
Surges, Line-to-line
IEC 61000-4-5
Conducted disturbance induced
by RF fields
IEC 61000-4-6
Rated
power
magnetic fields
IEC 61000-4-8
Voltage dips, short interruptions
and voltage variations
IEC 61000-4-11
Variation of power frequency
IEC 61000-4-28
All manuals and user guides at all-guides.com
IEC 60601-1-2:2014 test levels
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
IEC 60601-1-2:2014 Table 9
± 1 kV, ±2 kV
100 kHz repetition frequency
± 0,5 kV, ± 1 kV
230 VAC
6 Vrms
0.15 MHz – 80 MHz
80% AM at 1 kHz
frequency
30 A/m
50 Hz
0 % U
for 0,5 cycle at 0°, 45°,
T
90°, 135°, 180°, 225°, 270°
and 315° phase angles
0 % U
for 1 cycle
T
70 % U
for 25 cycles
T
0 % U
for 250 cycles
T
230 VAC
N/A
Additional test levels
± 6 kV contact
20 V/m
800 MHz – 2.5 GHz
80 % AM at 1 kHz
N/A
± 1 kV, ± 2 kV
5 kHz repetition frequency
± 0,5 kV, ± 1 kV
100 VAC and 240 VAC
N/A
30 A/m
50 Hz and 60 Hz
0 % U
for 1 cycle
T
40 % U
for 10 cycles
T
70 % U
for 25 cycles
T
100 VAC and 240 VAC
49 Hz and 51 Hz
100 VAC and 240 VAC, 50 Hz
59 Hz and 61 Hz
100 VAC and 240 VAC, 60 Hz
support.R82.org