Instant Help - KERN KFB-TAM Instructions D'utilisation Et D'installation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

11 Instant help

In case of an error in the program process, briefly turn off the display unit and
disconnect from power supply. The weighing process must then be restarted from the
beginning.
Help:
Fault
The displayed weight does
not glow.
The displayed weight is
permanently changing
The weighing result is
obviously incorrect
Should other error messages occur, switch display unit off and then on again. If the
error message remains inform manufacturer.
54
Possible cause
• The display unit is not switched on.
• Mains power supply interrupted (mains cable
defective).
• Power supply interrupted.
• (Rechargeable) batteries are inserted incorrectly or
empty
• No (rechargeable) batteries inserted.
• Draught/air movement
• Table/floor vibrations
• Weighing pan has contact with other objects.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
• The display of the balance is not at zero
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Warm-up time was ignored.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
KFB/KFN-TAM-BA_IA-e-1832

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Kfn-tam

Table des Matières