Overview Verified Weighing Systems (Contacts Of Circuit Board [K1] Short-Circuited By Means Of Jumper) - KERN KFB-TAM Instructions D'utilisation Et D'installation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

8.2 Overview verified weighing systems
(contacts of circuit board [K1] short-circuited by means of jumper)
In verified weighing systems the access to „P2 mode and „P4 tAr" is locked.
KERN KFB-TAM:
To disable the access lock, destroy the seal and actuate the adjustment switch.
Position of the adjustment switch see chap. 6.11.
KERN KFN-TAM:
In order to unlock the access, the seal must be destroyed and both contacts of the
printed circuit board [K2] must be short-circuited by a jumper, see chap. 6.11.
Attention:
After destruction of the seal the weighing system must be re-verified by an
authorised agency and a new verification wire/seal mark fitted before it can be
reused for applications subject to verification.
Menu block
Menu item
Main menu
Submenu
PO CHK
nEt H
Weighing with
tolerance range,
nEt LO
see chap. 7.7
PCS H
PCS L
BEEP
P1 COM
MODE
Interface
parameter
KFB/KFN-TAM-BA_IA-e-1832
Available settings / explanation
Upper limit value „Tolerance check weighing", input see
chap. 7.7.1
Lower limit value „Tolerance check weighing", input see
chap. 7.7.1
Upper limit value „Tolerance check counting", input see
chap. 7.7.2
Lower limit value „Tolerance check counting", input see
chap. 7.7.2
Acoustic signal for weighing with tolerance
no
range switched off
Audio sound when load is within tolerance
ok
limits
Audio sound when load is beyond tolerance
ng
limits
S0 off
CONT
S0 on
One output for stable weighing value
ST1
Continuous data output of stable weighing
STC
values
PR1
Output after pressing
Manual totalizing, see chap. 7.8
PR2
Press
added to the summation memory and issued.
AUTO
For automatic totalizing see chap. 7.9
This function is used to issue and add
individual weighing values automatically to the
summation memory on unloading of weighing
scale.
Continuous data output,
selectable "send zero" yes / no
and the weighing value will be
43

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Kfn-tam

Table des Matières