Télécharger Imprimer la page

Greenlee DM-500 Manuel D'instructions page 23

Publicité

Les langues disponibles

Les langues disponibles

Specifications
Display: 4-digit LCD (9999 maximum count) and 42-segment bar graph
Polarity: Automatic
Sampling Rate:
Numeric Display: 4 per second
Bar Graph Display: 20 per second
Temperature Coefficient: 0.15 x (accuracy) per °C below 18 °C or above 28 °C.
See Accuracy section for T1-T2 temperature coefficient.
Automatic Power-Off: After 17 minutes of inactivity
Noise Rejection*:
Normal Mode Rejection Ratio > 50 dB at 50 Hz and 60 Hz when measuring DCV
Common Mode Rejection Ratio > 60 dB from 0 Hz to 60 Hz when measuring ACV
Common Mode Rejection Ratio > 100 dB at 0 Hz, 50 Hz and 60 Hz when measuring DCV
Overload Protections:
Volts: 1000V peak, 780 VAC RMS
A: 15A/600V HBC F fuse, interrupting rating 100 kA, 13/32 x 1-1/2"
µA, mA and T2: 0.16A/250V F fuse, interrupting rating 1.5 kA, 5 x 20 mm
Other Functions: 600 VDC/VAC RMS
Overvoltage Categories:
+ Terminal: Category III, 600 VAC and 600 VDC; Category II, 750 VAC and 1000 VDC
µA and mA Terminal: Category III, 250 VAC; Category II, 250 VDC
A Terminal: Category III, 600 VAC; Category II, 250 VDC
Operating Conditions:
0 °C to 35 °C (32 °F to 95 °F), 0% to 80% relative humidity (non-condensing)
35 °C to 40 °C (95 °F to 104 °F), 0% to 70% relative humidity (non-condensing)
Altitude: 2000 m (6500') maximum
Indoor use only
Storage Conditions: –20 °C to 55 °C (–4 °F to 131 °F)
0% to 80% relative humidity (non-condensing)
Remove battery.
Battery: 9-Volt battery (NEDA 1604, JIS 006P or IEC 6F22)
Pollution Degree: 2
All manuals and user guides at all-guides.com
DM-500 • DM-510
23

Publicité

loading

Ce manuel est également adapté pour:

Dm-510