Er:1:11
... the display
– At least one weighing cell is not correctly connected or the connection is back to front.
Check the connections. The colours of plug and socket must match.
Switch the scale off using the Start key and restart. The scale then works normally
again. If this is not the case, inform Servicing.
Er:1:12
... the display
– The maximum weight for switching on was exceeded when the scale was switched on,
e.g. if bed and patient are being weighed. Remove the load from the scale and start it
again. The scale then works normally again. If this is not the case, inform Servicing.
... one segment is permanently illuminated or is not illuminated at all?
– The relevant segment is faulty. Inform Servicing.
8. Maintenance/verification
Have the scale verified by authorised persons in accordance with national legal require-
ments. Verification is essential if one or more security marks are damaged or the verifica-
tion counter does not match the number on the applicable verification counter mark (cf.
page 27).
We recommend having your scale serviced by your local service agent before having it
verified. seca Customer Service will be pleased to give you further assistance.
9. Technical data
Dimensions
width:
height:
depth:
weight
Measuring feet
width:
height:
depth:
weight:
Cable length:
Size of figures in display
Temperature range
Calibration as per Directive
2014/31/EU
Medical product as per Directive
93/42/EEC
Maximum load
Minimum load
Model 985
appears?
appears?
520 mm
927 mm
562 mm
approx. 25 kg
328 mm
150 mm
212 mm
approx. 6.2 kg
2.80 m
20 mm
+10 °C to +40 °C
verified, Class III
Class I with measuring function
250 kg
2 kg
GB