4. Click on
Quick Reference Table -- Agent Setup
Click on the following items to execute Agent setup functions:
Menu Item
Description
Agent Scale
Sets Agent waveform scale
Agent Display
Displays a separate Agent parameter box
Note: Ghosted if the monitor is displaying com-
bined MultiGas parameter box (page 15)
ID Override
Configures Scio module to identify an agent
automatically or to measure the concentration
levels of a user-specified anesthetic agent.
Note: The presence of organic cleaning solu-
tions or gases containing freon may impact the
accuracy of the Scio module.
Auto Zero Delay Delays automatic zeroing for 5 minutes for
uninterrupted monitoring.
Note: Gas sensors in the Scio module are
automatically zeroed and calibrated against
room air. During zeroing, the monitor tempo-
rarily blanks Scio parameter values. One
minute before automatic zeroing, the monitor
sounds an attention tone and displays the
message Auto zero in 1 minute.
Agent Alarms
Accesses Agent alarms in Alarm Limits table (see SC 7000/8000/9000XL
User's Guide Chapter 5)
Page 14
to display the Agent menu.
AGENT
The Agent Setup Menu
WARNING: Delaying the auto zero
may inpact the accuracy of the
device.
Scio MODULE D011100
Settings
• 1, 2, 3, 5, 10, or 20%
• ON, OFF
• OFF -- Module automati-
cally identifies agent and
displays its concentration
levels
Manual override selections:
• HAL
• ISO
• ENF
• SEV
• DES
Note: If you select an
agent from the ID Override
menu that is different from
the one actually adminis-
tered, the monitor displays
a message indicating the
type of agent it actually
detects (i.e. Hal detected).
N/A
SC 7/8/9XL VF2.x