• Firmly grip the body of the
handpiece;
CAUTION: Do not grab the
handpiece by the nozzle and/or by
the cord. Firmly hold the body of the
handpiece and rotate only the torque
wrench. Do not rotate the body during
tightening.
• Turn the torque wrench in a
clockwise direction until the notch
clicks (the external body of the
torque wrench turns with respect to
the body of the handpiece, emitting
mechanical sounds);
• The insert is now optimally locked;
The position of the LED light on the
nose cone of the MANIPOLO ABLATORE
LED (LED SCALER HANDPIECE - cod.
02150142) can be adjusted as follows:
• Grip the handpiece body and
slightly unscrew the metal ring
at the base of the nose cone by
turning it counter-clockwise;
• rotate the nose cone so that the
LED light is in the desired and
necessary position;
• to fix the position screw the metal
ring in a clockwise direction.
Select the desired power and irrigation
level based on the instructions provided
by the manufacturer of the dental unit.
NOTE: Activate the WATER function to
perform a treatment requiring irrigation,
and deactivate it to perform a treatment
that does not require irrigation.
Activate the COMPACT PIEZO LED.