Problem
Pump does not work
There is no power
Lip valve is missing, incorrectly
positioned or damaged
Pump is being held at an angle
Breastmilk container is full and lip
valve is already touching the
expressed breast milk
Poor pump suction
Low battery power
Suction strength is set too low
Pump suction too
Suction strength is set too high
powerful
Pump makes unusual
Lip valve cannot handle the flow of
noises
breast milk
Medicinal product to 93/42/EEC
NUK Vented Teat
Temperature range
- Operation
- Storage
Power supply:
- Mains operation via adapter
Type KSS07-045-0800
- Battery operation
All manuals and user guides at all-guides.com
12. Troubleshooting
Reason

13. Technical Data

Class I
Complies with EN 14350
8 °C to +40 °C
8 °C to +40 °C
Input 100 – 240 V~/50 Hz, output 4.5 VDC,
800 mA, hollow plug: polarity inner +, outer –
3 Type AAA batteries, 1.5 V
26
Remedy
For battery operation: check that the batteries
are correctly inserted.
For mains operation: check that the mains plug
or mains adapter is correctly inserted.
Check that the lip valve is correctly positioned. If
it is damaged use provided replacement lip valve.
Hold the pump upright. If the pump is held at an
angle it will switch off automatically.
Exchange breastmilk container.
Check whether there is sufficient battery power.
If a light is illuminated beneath the On/Off but-
ton the batteries are too weak. Insert new bat-
teries.
Turn the suction power regulator towards "+" to
attain the optimum suction strength.
Turn the suction power regulator towards "-".
Reduce the suction strength. With maximum
milk flow a lower suction strength is sufficient to
allow efficient expressing.