Guidance and manufacturer's declaration - electromagnetic emissions
The electrical breast pump NUK e-Motion is intended for use in the electromagnetic environment specified
bellow. The costumer or the user of the electrical breast pump NUK e-Motion should assure that it is used in
such an environment.
Immunity
IEC 60601-1 test
test
level
conducted RF
3 Veff
IEC 61000-4-6
150 kHz to 80 MHz
Radiated
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
NOTE: At 80 MHz and 800 MHz the separation distance for the higher frequency range applies.
NOTE : This guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Field strength from fixed transmitters, such as base stations for radio (Cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should
be considered. If the measured field strength in the location in which the electrical breast pump NUK e-Motion
is used exceeds the applicable RF compliance level above, additional measures may be necessary, such as
reorientation or relocating the electrical breast pump NUK e-Motion.
All manuals and user guides at all-guides.com
Compliance
Electromagnetic environment - guidance
level
3 V
Portable and mobile RF communications equipment
should be used no closer to any part of the Public
Access Defibrillator (COR 043), including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of the
transmitter.
Recommended Separation distance:
d= 1,17 * SQRT(P/W)
3 V/m
d= 4 * SQRT(P/W) for 80 MHz to 800 MHz
d= 7,67 * SQRT(P/W) for 800 MHz to 2,5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strength from fixed RF transmitters, as deter-
mined by an electromagnetic site survey, should be
less than the compliance level in each frequency
range.
Interference may occur in the vicinity of
equipment marked with the symbol "nonionising
radiation".