Cleaning And Storing The Device And Cuff - Beurer medical BC 58 Mode D'emploi

Masquer les pouces Voir aussi pour BC 58:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
In such cases, repeat the measurement. Ensure you do not
move or speak.
If necessary, re-insert or replace the batteries.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section "Technical specifica-
tions", the technical alarm will appear on the display indicat-
ing either " " or "Lo". In such cases, you should seek medical
assistance and check the accuracy of your procedure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard
IEC 60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear auto-
matically after about 8 seconds.

10. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under
water, as this can cause liquid to enter and damage the
device and cuff.
All manuals and user guides at all-guides.com
• If you store the device and cuff, do not place heavy
11. Specifications
Model no.
Measurement
method
Measurement
range
Display accuracy
Measurement
inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible operat-
ing conditions
26
objects on the device and cuff. Remove the batteries.
BC 58
Oscillometric, non-invasive blood
pressure measurement on the wrist
Cuff pressure 0 – 300 mmHg,
systolic 60 – 260 mmHg,
diastolic 40 – 199 mmHg,
Pulse 40 –180 beats/minute
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 90 mm x W 68 mm x H 30 mm
Approx. 158 g (without batteries)
140 to 195 mm
+ 5 °C to + 40 °C, ≤ 90 % relative air
humidity (non-condensing)

Publicité

Table des Matières
loading

Table des Matières