PREPARATION

1. Pull out the air hose from the nebulizer leaving it plugged to the air outlet pipe of the
device.
2. Rotate the upper part of the nebulizer anticlockwise.
3. Disconnect the internal pisper at the bottom of the nebulizer with finger force.
CLEANING
Clean all the components of the nebulizer (except the air hose) before and after each use by
choosing one of the methods described below.
Method 1: Thoroughly clean the components for 5 minutes by using hot drinking tap water
(around 40°C – 104°F) and neutral soap.
Method 2: Clean the components (except the air hose) by immersion in a solution of 60%
of water and 40% of white vinegar. At the end of the operation, rinse with plenty of hot
drinking water (around 40°C – 104°F).
At the end of the cleaning operations, rinse thoroughly by removing excess water and allow
to air-dry in a clean spot.
DO NOT BOIL OR AUTOCLAVE THE AIR HOSE AND THE MASKS
DO NOT WASH THE ACCESSORIES IN THE DISH WASHER
WASHING
Where diseases with risk of infection and microbial contamination are present, the end user
shall carry out the washing operations properly. The washing procedure can be performed
only if the components have been previously cleaned (see cleaning section).
For the washing procedure, the following operations shall be carried out:
- Fill a container of suitable size to hold every individual component with drinking water and
disinfectant (hypochlorite solution easily available in pharmacies) by complying with the
proportions indicated on the disinfectant packaging.
- The immersion time in the solution is indicated on the packaging of the hypochlorite solu-
tion according to the concentration chosen for the preparation of the solution.
- Rinse with plenty of warm drinking water until removing any trace of solution, dry and
store in a dry and dust-free place.
- Dispose the solution used according to the instructions provided by the manufacturer of
the disinfectant.

SUPPLIED ACCESSORIES

ACCESSORIES
HI-FLO accessories kit
(HI-FLO ampoule, Adult Mask, Children Mask,
Air hose, Mouthpiece and Nosepiece)
Air filter (+2 spare parts)
Transport bag
ENGLISH
20