The TIH 030m is equipped with the following safety features:
Automatic overheating protection
Automatic current control
Over-current circuit breaker.
In the TEMP MODE the heater will switch off if the temperature probe does not
register a temperature increase of 1°C (1.8°F) every 15 seconds (0.25 minute).
To increase the interval to 30 seconds (0.50 minute), press MODE and DOWN at the
same time.

7. Troubleshooting

A system fault will be indicated by an acoustic signal and one of the following fault
codes on the remote control display:
Display
Fault
E03 E
Overheated coil
E05 E
Temperature increase of
less than 1°C (1.8°F)
every 15 seconds
(or 1° every 30 seconds)
E06 E
Temperature probe not
connected (or defective)
or excessive temperature
drop
E10E
Electronics communication
problem
E11E
Electronics communication
problem
E12E
Electronics communication
problem
12
Action
Wait until the inductive coil cools
Switch the heater OFF and then back ON
Check the temperature probe connection
If the connection is OK, select the 30 second
interval as described in section 6 or operate
the heater in TIME MODE
Check the temperature probe
Switch the heater OFF and then back ON
If problem remains return the TIH to SKF
for repair
Switch the heater OFF and then back ON
If problem remains return the TIH to SKF
for repair
Switch the heater OFF and then back ON
If problem remains return the TIH to SKF
for repair