CSA Control Drawing

Conductivity Measuring Loop
Terminals
1, 2, 3, 4, 5, 6
IIC (GRP A, B)
IIB (GRP C)
IIC (GRP D)
Temperature Measuring Loop
Terminals
7, 8
IIC (GRP A, B)
IIB (GRP C)
IIC (GRP D)
All Combined Outputs
Terminals
1, 2, 3, 4, 5, 6, 7, 8
IIC (GRP A, B)
IIB (GRP C)
IIC (GRP D)
96
Uo, Vsc Io, Isc
Po
9V
111mA
139mW
9V
111mA
139mW
9V
111mA
139mW
Uo, Vsc Io, Isc
Po
5V
3.5mA
5mW
5V
3.5mA
5mW
5V
3.5mA
5mW
Uo, Vsc Io, Isc
Po
9V
114.5mA
144mW
9V
114.5mA
144mW
9V
114.5mA
144mW
Transmetteur Cond Ind 7100e/2(X)H
Co, Ca
Lo, La
4µF
1.5mH
12µF
6mH
32µF
12mH
Co, Ca
Lo, La
100µF
1H
300µF
1H
800µF
1H
Co, Ca
Lo, La
4µF
1.5mH
12µF
6mH
32µF
12mH