TENZ2174T
parameter. After elapsing this period, the current value is
measured and used for calculation of new calibration constants
and all data are stored in the memory. This action is indicated with
"SAUE" displaying. The device then continues to the next
parameter "CALH". If the "Menu" pushbutton is depressed for the
period exceeding 5 seconds during "dELA" interval countdown,
then neither measuring nor calculation and storage in the memory
takes place and the name of the next parameter is displayed.
In case of "CALH" parameter, the setting procedure is similar,
however, the sensor is loaded with the nominal value of the signal
measured. New calibration constants are always calculated and
stored in the memory for both calibration parameters.
If you need to perform measuring without calibration
calculation, depress one of the pushbuttons at power supply
turning on. "AdHo" value will be displayed and measuring with
direct displaying of the measured values takes place after
pushbutton releasing.
4. Technical specification
Power supply:
Current consumption:
Relay output contact:
Displaying range:
Box dimensions:
Device protection:
Tensometer power supply:
Operational temperature:
Operating Instructions
24V DC
max. 60mA
max. 60VDC/125VAC/ 1A
-9999 up to 9999
70 x 90 x 65 mm (w x h x d)
IP20
5V DC
5 to 40 °C
14