7.
To Stop the bleaching device at any time press the START / STOP button. If the device is not
manually stopped it will stop after the selected time has elapsed. Note that the fans will
continue to work for some time in order to cool down the unit.
8.
At the end of the day turn off the device by holding down
the Start / Stop button for 3-4 seconds or by disconnecting it
from the power network.
9. In case of overheating the thermal protection is triggered
and the device will cease to emit light.
The display will indicate "Oh" and the fans will turn on
for 1 minute.
After cooling, the device will continue its normal operation.
Overheating can only happen in an emergency situation
or by accidental clogging of the cooling vents.
That's why the cooling vents located on the bottom
of the bleaching head must always be unobstructed
in order to allow air circulation.
VII. DAILY CARE AND MAINTENANCE
1.
Disinfect the PVC window on the bleaching head before
and amer each pa>ent with cojon or other nonabrasive
cloth soaked in alcohol.
2.
To disinfect the device and its parts, spray disinfectant
onto a piece of som cloth or cojon and clean it.
Do not use abrasives or solvents as this may cause damage.
A
O
E
O
CLICK
ARCUS
LIPS
ARCUS
DUETTO
CARE
LED Bleaching Lamp - MOBILE