Technical Data - Tintometer Lovibond MD100 Mode D'emploi

Photometer-system
Masquer les pouces Voir aussi pour Lovibond MD100:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 72

Technical Data

GB
Instrument
Light source:
Wavelength accuracy
Photometric accuracy*
Photometric resolution
Power supply
Operating time
Auto-OFF
Display
Storage
Serial Interface
Time
Calibration
Dimensions
Weight
Ambient conditions
Waterproof
CE
*measured with standard solutions
To ensure maximum accuracy of test results, always use the reagent systems
supplied by the instrument manufacturer.
70
Technical Data
three
wavelength, automatic wavelength selection,
direct reading colorimeter
LEDs,
interference filters (IF) and photosensor in transparent
cell chamber. Wavelength specifications of the IF:
430 nm
= 5 nm
530 nm
= 5 nm
560 nm
= 5 nm
± 1 nm
3% FS (T = 20° C – 25° C)
0.01 A
4 batteries (AAA/ LR 03)
17hr operating time or 5000 test measurements in
continuous mode when display backlight is off
automatic switch off
20 minutes after last keypress
backlit LCD (on keypress)
internal ring memory for 16 data sets
IR interface for data transfer
real time clock und date
user and factory calibration
resetting to factory calibration possible
155 x 75 x 35 mm (LxWxH)
approx. 260 g (incl. batteries)
temperature: 5 – 40 °C
rel. humidity: 30– 90 % (non-condensing)
floating; as defined in IP 68 (1 hour at 0.1 meter)
Certificate for Declaration of CE-Conformity
at www.lovibond.com
MD100_5a 08/2016

Publicité

Table des Matières
loading

Table des Matières