The value under this mode is object surface
temperature instead of core temperature.
Exceeding measurement range
Body mode:
-
When measurement value is lower than 32.0°C, it
displays Lo (fig.9.10) and gives 4 beeps.
-
When measurement value is higher than 42.9°C, it
displays Hi (fig.9.11) and gives 4 short beeps.
Object mode:
-
When measurement value is lower than 0°C, it
displays Lo (fig.9.12) and gives 4 beeps.
-
When measurement value is higher than 100.0°C,
it displays Hi (fig.9.13) and gives 4 short beeps.
ATTENTION: When surrounding temperature is lower
than 10.0°C or higher than 40.0°C, it displays Err. It is not
allowed to measure, or accuracy is not assured.
Battery replacement
Open the battery lid and take out exhausted
battery.