Télécharger Imprimer la page

3B SCIENTIFIC PHYSICS U8531401 Instructions D'utilisation page 7

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 9
3. Technical data
Ranges (with reference to 1 V output voltage):
Voltage:
Current:
Charge:
Input resistance:
Voltage:
Current/charge:
Precision:
Measurement output:
Output voltage:
Operating voltage:
Dimensions:
Weight:
4. Operation
4.1 Voltage measurements
Caution:
Voltages in excess of 300 V may only be applied to
terminal "U" (10) with a measuring range of 100 V/1 V.
Connect voltmeter to terminal pair (5).
Calibrate the zero point for the voltmeter using the
"Offset" knob (7) where necessary (only relevant for
highly sensitive ranges).
Connect the voltage to be measured via a BNC cable
to socket "U" (10) and ground (11).
Set the polarity switch (6) to the right for positive
voltages or to the left of the measured voltage is
negative.
Range
Gain
100 V
0.01
10 V
0.1
1 V
1
100 mV
10
10 mV
100
1 mV
1000
100 μV
10000
Elwe Didactic GmbH • Steinfelsstr. 6 • 08248 Klingenthal • Germany •
3B Scientific GmbH • Rudorffweg 8 • 21031 Hamburg • Germany •
0.1 mV – 100 V (7 ranges)
-11
-4
10
– 10
A (7 ranges)
-11
-7
10
– 10
As (5 ranges)
10 MΩ
0 Ω, for voltages up to 300 V
3%
0 – 10 V
Zero point calibration and
reversible polarity
±15 V or 30 V, max. 50 mA
See reverse of apparatus
235 × 230 × 180 mm
2.8 kg approx.
Value corresponding to scale
limit of display instrument
10 V
3 V
1000 V
300 V
100 V
100 V
30 V
10 V
3 V
1 V
300 mV
100 mV
100 mV
30 mV
10 mV
10 mV
3 mV
300 μV
100 μV
1 mV
Subject to technical amendments
© Copyright 2008 3B Scientific GmbH
4.2 Current measurements
Connect voltmeter to terminal pair (5).
Calibrate the zero point for the voltmeter using the
"Offset" knob (7) where necessary (only relevant for
highly sensitive ranges).
Connect the current to be measured via a BNC
cable to socket "I Q" (9) and ground (11).
An output of 1 V to the display from the amplifier cor-
responds to the current value selected for the range.
Measured current = range value on front panel x de-
flection of voltmeter.
4.3 Charge measurements
Connect voltmeter to terminal pair (5).
Press "Discharge" button (12) to discharge any exist-
ing charge.
Conductors in the circuit may lead to inaccuracies
in the measurement (due to innate capacitance,
magnetic leakage). For this reason it is practical to
plug a contact pin directly into terminal (9) to mea-
sure the charge
The voltage displayed by the voltmeter corresponds to
the value of the charge. Here too, an output of 1 V to
the display instrument corresponds to the charge given
by the range setting of the amplifier.
Since the displayed value for a change is held static,
charges can be applied repeatedly. The value displayed
then corresponds to the sum of all the individual char-
ges. Any charge applied to the measuring amplifier can
be cancelled out by another charge of opposing sign.
4.4 Use of the measuring amplifier as a DC power
supply
To output a fixed DC voltage of 15 V:
Connect the positive pole of the load to either ter-
minal "L+" or "L-" (2 or 4). Connect the negative po-
le to terminal "0" (3).
To output a fixed DC voltage of 30 V:
Connect the positive pole of the load to terminal
"L+" (2) and the negative to terminal "L-" (4).
1 V
10 V
1 V
1 mV
directly.
www.elwedidactic.com
www.3bscientific.com

Publicité

loading