Wile 200 is equipped with error messages to guide the user to get the most reliable re-
sults, as well as to ensure that the meter is in good condition.
Error code
Critical Error
7
Weight not detected
8
Taring result is too high
Weight measurement uncertain /
9
meter shook during the measure-
ment
Moisture measurement result
10
stabilize slowly
Weight measurement too un-
11
certain / meter shook during the
measurement
Temperature of the sample lower
12
than 2 C / Measurement uncer-
tainty is high
Temperature of the sample above
13
50 C / Measurement uncertainty
is high
Large temp. diff. between the grain
14
and test cell / Measurement uncer-
tainty has increased
Not on a suffciently level surface,
15
meter warns with sound alarm
16
Weight result is over 330 g
17
Weigh result is under 50 g
Hectoliter weight is 20% higher
18
than average
Hectoliter weight is 20% lower
19
than average
Test cell is not cleared before the
20
measurement
Error Codes
Cause
15
Solution
The meter must be submitted to an
authorized service center
Repeat test / Do weight test
Do weight test and check the scale
Keep the meter stable,
measurement uncertainty has increased
Happens to very wet samples
Keep the meter stable, measurement
uncertainty has increased
The tester is most accurate when grain
and tester temperature is between 16C
and 32C
The tester is most accurate when grain
and tester temperature is between 16C
and 32C
For better accuracy, grain and meter
test cell should be at similar tempera-
tures
Keep meter stable during measurement
Check that leveling handle did not stay
above the test cell/ Do weight test
Check that leveling handle did not stay
above the test cell/ Do weight test
Measurement uncertainty has in-
creased, Do weight test
Measurement uncertainty has in-
creased, Do weight test
Clear test cell