Performing Measurements (cont'd)
Note:
If the instrument detects a faulty state during the auto-calibration, the fail icon (X) will
be displayed:
A common reason for failure is improper environmental conditions. This includes excessive hu-
midity or excessive temperature. In this case it is possible to perform measurements by pressing
the START/STOP button again but results could be outside of the technical specification.
Mains powered instrument operation
Connecting the instrument to the mains supply while the instrument is OFF will charge the bat-
teries. The instrument will remain OFF during this period. If the batteries are charging, a flashing
battery indicator will appear in the upper left corner of the LCD.
Note:
If batteries are defective or missing, the charger will not work.
Connecting the instrument to the mains supply while the instrument is ON will cause the instru-
ment to be powered by the mains supply. If the instrument is not in Insulation Measuring Mode*,
the internal charger will charge the batteries. If the batteries are charging, a flashing battery
indicator will appear in the upper left corner of the LCD.
Note:
It is not recommended to connect or disconnect the instrument to mains supply while the
instrument is in Insulation Measuring Mode*.
*Insulation Measuring Mode: When the instrument performing insulation measurements.
Backlight operation (battery powered instrument)
After turning the instrument ON, the LCD backlight is automatically turned ON. It can be turned
OFF and ON by pushing the
Backlight operation (mains powered instrument)
After turning the instrument ON, the LCD backlight is automatically turned OFF. It can be turned
OFF and ON by pushing the
Off function
The instrument can be switched OFF by pressing the ON/OFF button.
button.
button.
5990A
13