Télécharger Imprimer la page

B.A. International Optima OS600L Mode D'emploi page 11

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 32
5.3 Memory function
5.3.1 Memory button
Press the memory button to access memory address where detailed figures (Gear ratio, Torque, Speed, For/Rev, Coolant) of each function in program are saved.
Memory address cycles through numbers 1 to 9 upon pressing Memory button each time.
Memory
Button
The memory
number(1-9) currently
selected is being
displayed.
※ Initial setting (Factory settings) : Memory
Program
Drilling
Tapping
Remove tap
Implant
Remove
Rock screw
5.3.2 Saving data
Save detailed figures in memory which are currently set (Gear ratio, Torque, Speed, For/Rev, Coolant) for each function in program.
press and hold the memory button for 2sec to start saving data. A beep will sound and 'No.' will continuously flash.
Press the Memory button again to select the memory address where data is to be saved.
Press the memory button for 2sec once more to save the data. 'No.' will stop flashing and 2 beeps will sound again when saving is completed
Memory
Button
Memory 1
① Drilling set value
② Tapping set value
③ Remove tap set value
④ Implant set value
⑤ Remove set value
⑥ Rock screw set value
Gear ratio
Torque [Ncm]
Speed [rpm]
20:1
55
20:1
40
20:1
40
20:1
40
20:1
55
20:1
10
The memory number (1~9) currently selected is being displayed.
'No.' flashes continuously until data is saved.
...
Memory 9
...
① Drilling set value
...
② Tapping set value
...
③ Remove tap set value
...
④ Implant set value
...
⑤ Remove set value
...
⑥ Rock screw set value
Motor direction
1,500
Forward
50
Forward
50
Reverse
50
Forward
50
Reverse
50
Forward
10
Coolant level
4
3
3
3
0
0

Publicité