CLEANING
Proper care and cleaning is essential to ensure a long life of accurate and effective operation. Disconnect the scale from the battery charger.
1. Clean all external surfaces with a clean damp cloth or tissue. Mild soap and water solution may be used. Dry with a clean soft cloth.
2. Do not immerse the scale into cleaning or other liquid solution.
3. Do not use Isopropyl Alcohol or other solutions to clean the display surface.

TROUBLESHOOTING

Refer to the following instructions to check and correct any failure before contacting service personnel.
SYMPTOM
Scale does not turn on
Questionable weight or the scale
does not zero
The display shows "- 0F- "

CALIBRATION PROCEDURE

This scale has been calibrated at the factory and does not require re-calibration prior to use.
The following procedure should be performed only to correct any error in calibration. For maximum accuracy, calibration should be done using
certified weights. The recommended calibration weight is 600 lbs, however calibration can be performed using any weight no less than 44 lbs and no
more than 600 lbs.
1. Turn the scale on.
2. At the same time press the TARE/ZERO and KG/LB buttons. "t" and the number representing the current automatic shut off value will
appear on the display.
3. Press the TARE/ZERO button. "CAL" and then '600.0' will appear on the display.
4. To change the calibration weight, press the HOLD/RELEASE button to move the flashing indicator to the left. Then press the KG/LB button to set the
value to the desired calibration weight.
5. Press the TARE/ZERO button again. The display will show 'CAL 00'.
6. Press the TARE/ZERO button again. The display will show CAL-SP. Load 600 LB (or altered value) of calibration weight on to the scale and press the
TARE/ZERO button. The display will show "600.0" (or altered value).
7. Remove the weight from the scale. The scale returns to the normal weighing mode.
ERROR LIST
ERROR NO.
Err 1
A / D value is lower when CAL
--OF--
The load on the scale exceeds the capacity (600 lb / 270kg). Remove excess weight and use the scale according to its
Or-Err
High saturation state (High A / D)
Or-Err
Low saturation state (Low A / D)
UF
Battery is low.
POSSIBLE CAUSE
1. Dead battery
2. Faulty electrical outlet
3. Bad battery charger
1. External object interfering with the scale.
2. The display did not show "0.0" before
weighing
3. Scale is not on a level floor
4. Scale is out of calibration
The load on the scale exceeds the maximum
capacity (600 lbs / 270 kg)
DEFINITION
CORRECTIVE ACTION
1. Attach battery charger to recharge battery.
2. Use a different outlet for the battery charger.
3. Replace battery charger.
1. Remove interfering object from the scale.
2. Ask the patient to get off the scale and begin weighing
process again.
3. Place the scale on a level floor and begin weighing process
again.
4. Check weight with known weight value.
Remove the excess weight and use the scale according to its
limits.
CHECK / REPLACE
Scale is below minimum calibration weight. Calibration weight
must be more than 44 lbs. The recommended calibration weight
is 600 lbs.
limits.
The load cell cable is not connected properly. Check cables
and mechanical connection. If problem persists, contact
Health o meter Professional Customer Service 800-815-
6615.
The load cell cable is not connected properly. Check cables
and mechanical connection. If problem persists, contact
Health o meter Professional Customer Service 800-815-
6615.
Recharge battery. Overnight charge is required to fully charge
battery.