7.3 Foot switch settings

After connecting the foot switch at the back, the required mode must be set at the front. To set the
mode, press the button of the foot switch, until the LED is lit next to the desired setting.
Double and single footswitches can be used:
With the double foot switch you can activate the cutting or coagulation power (if set) by pressing
the pedal
With the single footswitch you can only activate the set mode.It must be selected by pressing the
foot switch symbol.

7.4 Handles with buttons

Activating the handle with buttons takes place independently of the foot switch settings. These cannot be
activated using the foot switch.

7.5 Neutral electrode settings and attachment

In the monopolar operating mode, a neutral electrode is required.
After attaching the neutral electrode to the patient, the HF unit recognizes the neutral electrode
automatically (cf. illustration on the left and chapter 4.2 Overview of symbols for operating modes and
settings). The HBS 100 recognizes neutral electrodes with divided and undivided contact surfaces.
!
If the displayed neutral electrode type does not agree with the used
neutral electrode type, this is indicative of poor adhesion of the
neutral electrode on the patient, a defective neutral electrode or a
Please note the following information when attaching the
defective cable!
neutral electrode
EN
43