WIKA DIH10-Ex Mode D'emploi page 19

Tête de raccordement avec module d'affichage numérique intégré
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5. Commissioning, operation
5.3.2 Programming the display
1. Connect instrument in accordance with connection diagram.
2. Switch on the current-loop supply (current between 4 ... 20 mA).
An initialisation and segment test is carried out. Then DIH10-Ex and
subsequently the firmware version number (e.g. F1.24) is displayed.
Following this, the display switches into operating mode.
3. Press the program button P, the program number P 0 will be
displayed.
4. By pressing P and the
change the program number.
5. Under the selected program number, switch to the stored value with
the P button.
6. By pressing P briefly, the required digit can be selected. The selected
digit is changed using
7. The entry is saved by pressing the P button for approx. 2 seconds; the
value has been saved once the display shows horizontal bars.
8. If no button is pressed, after approx. 7 seconds the instrument returns
to operating mode.
Additional functions in normal operation to retrieve the recorded
Min/Max values
Using the
in the display for some seconds.
Using the
in the display for some seconds.
By simultaneously pressing the
values are deleted.
If the measuring range setting is changed, to maintain the measurement
accuracy, the display must be recalibrated.
WIKA operating instructions model DIH10-Ex
button or P and
or
button, the recorded value of the Max memory is shown
button, the recorded value of the Min memory is shown
button simultaneously,
.
and
buttons, the Min and Max
EN
19

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Dih10

Table des Matières