Do NOT use to mask or relieve
undiagnosed pain. This may delay
diagnosis of a progressive condition.
Do NOT use if you have, in the area
being treated: active or suspected
cancer or undiagnosed pain with a
history of cancer. Stimulation directly
through a confirmed or suspected
malignancy should be avoided as it may
stimulate growth and promote spread of
cancer cells.
Do NOT use electrodes on the front of
the neck. Stimulation on the front of the
neck can affect your heart rate or cause
contraction of the throat.
Do NOT use electrodes across the
chest. Very strong stimulation across
the chest may cause an extra heartbeat
and/or rhythm disturbances to your
heart, which could be lethal.
Do NOT use Perfect TENS while
simultaneously
connected
frequency surgical equipment as it may
result in burns at the site of stimulator
electrodes and possible damage to the
stimulator.
Do NOT use Perfect TENS in close
proximity (e.g. 1 m) to a shortwave or
microwave
as
this
instability in the stimulator output.
Do NOT use electrodes near the thorax
as this may increase the risk of cardiac
fibrillation.
CAUTIONS:
Caution should be used if you have a
bleeding disorder as stimulation may
increase blood flow to the stimulated
region.
Caution should be used if you have
suspected or diagnosed epilepsy as
to
high
may
produce
8
electrical stimulation may affect seizure
threshold.
Caution should be observed when
using the device at the same time as
being
connected
equipment with body worn electrode
pads. It may interfere with the signals
being monitored.
Caution should be used following
recent surgical procedures. Stimulation
may disrupt the healing process.
Caution Not intended for use in an
oxygen rich environment.
Caution Not intended for use in
conjunction
with
anaesthetics or flammable agents.
Caution The patient is an intended
operator.
Caution Do not service and maintain
the device while in use.
Caution Maintenance and all repairs
should only be carried out by an
authorized agency. The manufacturer
will not be held responsible for the
results of maintenance or repairs by
unauthorized persons.
If necessary, we will provide circuit
diagrams, component part lists or other
information that will assist authorized
service personnel to repair the device.
Caution The operator should not touch
the patient at the same time when
touching the battery output.
Caution Keep away from pets and
pests.
Caution Do not permit use by children
unable to understand the instructions or
persons with cognitive disabilities, i.e.;
Alzheimer's disease or dementia.
Caution Keep away from children
under 5 years of age. Long cord - risk
of strangulation in infants.