Battery Operation - Adam Equipment SHS Serie Mode D'emploi

Masquer les pouces Voir aussi pour SHS Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN 
 Clean and protect your scale after use. 
Overload Warning 
Please  do  not  add  any  item  that  exceeds  the  maximum  capacity  of  the  scale. 
When the display shows "- - - - - " remove the item to avoid damage to the load 
cell and scale. 
 
3.2 BATTERY
OPERATION 
 
 
 The  lithium  battery  will  operate  continuously  for  20  hours  before 
requiring recharging. The scale will alert you that the battery is low on 
charge when the Indicator LED above the battery symbol lights up. 
 To charge the battery, unclip the battery cover on the rear of the scale 
and  insert  the  connector  into  the  socket  as  marked.  Plug  the  charger 
into  the  mains  power  supply  and  wait  until  the  LED  on  the  charger 
shows green. The battery is now fully charged. 
 Charging time for a completely discharged battery is approximately 16 
hours.  A  full  battery  pack  is  recommended  to  keep  the  scale 
continuously operational. 
 Only use the charger as supplied with the unit. 
 The switch inside the battery cover must be in position A to make use 
of the lithium battery. 
 If using AA or NiMH  batteries the switch must be in position B 
 The scale will automatically power down after 30 minutes of non use to 
further protect the battery. 
 To replace the lithium battery pack please contact your dealer 
 
 
 
© Adam Equipment Company 2012
7