This section includes a list of error messages and frequently asked questions for
problems you may encounter with your blood pressure monitor. If the monitor is
not operating as you think it should, check here before arranging for servicing.
LCD Display
Information
Err is displayed in the
SYS area, 1 is displayed
in the DIA area
Err is displayed in the high
SYS area, 2 is displayed
in the DIA area.
Err is displayed in the
SYS area, 3 is displayed
in the DIA area.
Err is displayed in the
SYS area, P is displayed
in the DIA area.
HI is displayed in the DIA
area.
Troubleshooting
Reason / Meaning
The leakage is too fast or the pulse signal is too
weak. Please check the cuff, re-fasten it around
your arm and try again.
The blood pressure signal cannot be detected
due to too much noise. Please remove the noise
sources and measure again.
The test result is abnormal. Please measure again.
The inflation fails. Please check the cuff, re-fasten it
around your arm and try again.
The inflation pressure is greater than 307±8 mmHg
( 41 kPa±1 kPa ). Please measure again.
165