Xylem SI Analytics TitroLine 7500 KF Mode D'emploi page 112

Titrateur
Masquer les pouces Voir aussi pour SI Analytics TitroLine 7500 KF:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

112
Fig. 72
Please confirm your input with <OK</<ENTER>.
4.6.2.1 Calculation Formula
The appropriate calculation formula is selected on the Formula selection submenu:
Fig. 73 a
The following calculation formulae are available for automatic titration mode:
Formula
EP
(EP-B)*T*M*F1/(W*F2)
(W*F2)/(EP-B)*M*F1)
The abbreviations used here have the following meaning:
EP:
Consumption at the end point in ml
B:
Blank value in ml. Mostly determined by way of titration
T:
Titer of the titration solution (e.g. 5.012)
M:
Mol; mol- or equivalence weight of the sample
F1-F5
Factor 1-5. conversion factors
W
"Weight", weighed-in quantity in g or volume in ml."
Additional information
Formula for calculating only the ml
consumption
Formula for calculating the
concentration of a sample taking into
account a blank value in terms of ml.
Formula for calculating a titer (T) of a
titration solution.

Publicité

Table des Matières
loading

Table des Matières