ARM ADJUSTMENT

The LS200 Procedure Light has been
designed to allow the user to adjust the
amount of resistance at the joints of the arm.
The main joint features a large knob for
easy adjustment of this key joint. Turn knob
to achieve required tension.
NOTE: There is no need to adjust knob
for every use. The set tension will be
maintained over multiple movements.
NEVER completely unscrew large knob.
The top and lower joints have been preadjusted at Welch Allyn.
Should further adjustment be necessary, the joint resistance can
be increased or decreased by using a screwdriver to adjust the
set screws at point (A) or (B).

LAMP REPLACEMENT

WARNING: The lamps operate at high temperatures. DO NOT
attempt to replace lamps while they are hot. Wait until the lamps
are cool. Lamps may shatter under pressure, handle with care.
Replace with Welch Allyn lamp #06400 only.
1. Turn power switch to OFF position and unplug the power cord
from the electrical outlet.
2. Rotate luminaire so that the bezel and handle are facing
you (see A).
3. Turn handle counterclockwise to unscrew (see B). Remove
handle and carefully slide bezel out and allow to rest on wire
tether (see C).
4. Locate release lever on lamp holder and push downward (see D).
BEZEL
HANDLE
A
B
9
A
ADJUSTMENT
B
KNOB
TURN
COUNTER-
CLOCK-
WISE