TRANSFORMER MOUNTING

1. Select proper mounting hardware from hardware kit in the
LS200 mounting option box: Choose 1" (2.5cm) wood screws
for mounting into wood stud and 1" (2.5cm) sheet metal screws
and plastic anchors for concrete or brick.
2. Select a location for the transformer that is within 8' (2.4m) of an
electrical outlet and within 1.5' (0.5m) of the wall mount plate to
allow enough slack for full range of arm motion.
3-1/8"
3. Using the template indicators on
8cm
left side of this page (3-1/8"
[8cm] on center), mark the location
for drill holes on the wall. Make
sure they are vertical.
For wooden walls:
No pilot hole is needed. Use
1.5" (3.8cm) wood screws
from transformer hardware kit.
For concrete walls:
Using a hammer drill and
3/16" (2.3mm) masonry
drill bit, drill holes for plastic
anchors to a minimum of
1" (2.5cm) deep. Tap in
plastic anchors from trans-
former hardware kit until flush.
4. Drive screws into the wall, allowing screw heads to protrude
11/32" (2.3mm).
5. Mount transformer by fitting round holes in the back of trans-
former over mounting screws on wall, then slide downward
until secure.
6. Attach power cord to transformer.

OPERATING INSTRUCTIONS

TRANSPORT INSTRUCTIONS
FOR MOBILE STAND UNIT
1. To prevent the light from tipping, position
arm at a 45˚ angle or greater, when rolling
the LS200 to a new place.
2. Center arm in range of motion.
3. Unlock casters by lifting up on the tabs and
hold unit at top of pole while moving it.