3 Reprocessing methods
Recommended sterilization methods or parameters are as follows:
Pressure impulse method
S cycle
Vacuum method B cycle
* Depending on the type of use. National guidelines have to be observed.
 Follow the instructions of the autoclave manufacturer regarding the sterilization of this medical products for autoclaves with S cycle
method.
Pursuant to U.S. and Canadian standards:
Dynamic Air Removal:
Unwrapped
Dynamic Air Removal:
Wrapped or pouched
NOTE! This table represents the variation in sterilizer manufacturers' recommendations for exposure at different temperatures!
 For a specific sterilizer, consult only that manufacturer's recommendations.
Page 20
134°C (273.2°F)
Minimum pressure 300 kPa
134°C (273.2°F)
Minimum pressure 300 kPa
135°C (275°F)
Minimum sterilization time = 3 minutes
135°C (275°F)
Minimum sterilization time = 3 minutes
Minimum
3.5 - 5 - 18 minutes*
sterilization time
Minimum
3.5 - 5 - 18 minutes*
sterilization time
Minimum
15 minutes
drying time
Minimum
10 minutes
drying time
Drying time not applicable
Drying time = 60 minutes
STATIS