Acelity KCI PREVENA PLUS Mode D'emploi page 36

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Table 205
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Test
Band
a
Frequency
(MHz)
(MHz)
385
380 - 390
450
430 - 470
710
745
704 - 787
780
810
870
800 - 960
930
1 720
1 845
1 700 -
1 990
1 970
2 450
2 400 -
2 570
5 240
5 100 -
5 500
5 800
5 785
NOTE: If necessary to achieve the Immunity Test Level, the distance between the transmitting antenna and the ME
Equipment or ME System may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a
For some services, only the uplink frequencies are included.
b
The carrier shall be modulated using a 50% duty cycle square wave signal.
As an alternative to FM modulation, a 50% pulse modulation at 18 Hz may be used because while it does not
c
represent actual modulation, it would be worst case.
Service
Modulation
a
Pulse modulation
TETRA 400
18 Hz
GMRS 460,
FM
±5 kHz deviation
c
FRS 460
1 kHz sine
LTE Band
Pulse modulation
13, 17
217 Hz
GSM 800/900,
TETRA 800,
Pulse modulation
IDEN 820,
18 Hz
CDMA 850,
LTE Band 5
GSM 1800;
CDMA 1900;
GSM 1900;
Pulse modulation
DECT;
217 Hz
LTE Band 1, 3
4, 25; UMTS
Bluetooth
WLAN,
Pulse modulation
802.11 b/g/n,
217 Hz
RFID 2450,
LTE Band 7
WLAN 802.11
Pulse modulation
a/n
217 Hz
Maximum
b
Power
(W)
b
1,8
2
b
0,2
b
2
b
2
b
2
b
0,2
36
Distance
Immunity Test
(m)
Level
(V/m)
0,3
27
0,3
28
0,3
9
0,3
28
0,3
28
0,3
28
0,3
9

Publicité

Table des Matières
loading

Produits Connexes pour Acelity KCI PREVENA PLUS

Table des Matières