Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

BEGINNING THERAPY

Blockage Alert
Alert Mute
Button
1. Ensure the PREVENA PLUS™ Dressing has been applied as described in the PREVENA PLUS™
CUSTOMIZABLE™ Dressing Application section.
NOTE: To interrupt therapy or turn unit off, press and hold center of the On/Off button for three
seconds.
Once therapy is on for one hour non-stop, the seven day lifespan begins and continues even if
unit is turned off.
3. With therapy on, assess dressing to ensure integrity of seal.
4. If there is any evidence of a leak, check the SENSAT.R.A.C.™ Pad and the dressing seals, tubing
connectors, and canister connection, and ensure clamp is open. Refer to the Correcting a Leak
Condition section for more information.
On/Off
Therapy Life
Button
Indicator
2. To begin therapy, press and hold center of the On/Off button for
three seconds. The PREVENA PLUS™ 125 Therapy Unit, while in
operation, may have a moderate sound emanating from the unit.
All seven Therapy Life Indicators will illuminate with a green LED,
indicating therapy is running.
The dressing should have a wrinkled appearance and the foam
3
bolster should be compressed.
If the foam bolster is not compressed or the therapy unit alerts,
X
see the Alerts section.
Leak Alert
Power Connected
Battery Level
20

Publicité

Table des Matières
loading

Table des Matières