SECTION 8—TROUBLESHOOTING
GUIDE
SYMPTOM:
Alarm:
Short beeps, long
pause
Concentrator not
operating,
power switch On.
Beep......Beep......
Alarm:
Continuous
Concentrator not
operating, Power
Switch On.
Beeeeeeeep....
Part No 1163145
SECTION 8—TROUBLESHOOTING GUIDE
PROBABLE CAUSE:
Main Power Loss:
1. Power cord not
plugged in.
2. No power at outlet.
3. Tripped circuit
breaker.
System Failure:
1. Unit overheating due
to blocked air intake.
2. Insufficient power at
outlet.
3. Internal repairs
required.
1. Insert plug into outlet.
2. Inspect house circuit
breakers or fuses. If problem
recurs, use a different outlet.
3. Push/reset circuit breaker.
If problem recurs, call Service
Provider.
1a. Remove and clean cabinet
filters.
1b. Move oxygen
concentrator at least twelve
inches away from walls,
draperies or furniture.
2. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
3. Call Service Provider.
29
SOLUTION:
Perfecto
™Series
2